PM1214 NUANCE SR RF; PM1230 and PM2230 NUANCE MRI; PM2214 NUANCE DR RF These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.
FDA Recall Z-0034-2018 · Posted None
Advertisement
Recalling FirmSt Jude Medical Inc.
Recall StatusTerminated
Product CodeNKE
Date Initiated2017-08-28
Date Terminated2020-07-08
Product Quantity2741 units
Reason for Recall
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
Firm Action
Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
Worldwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.