TxRx 15CH Knee Coil 3T QED part number: Q7000051 Siemens Model # 10606525
FDA Recall Z-0035-2020 · Posted None
Advertisement
Recalling FirmQuality Electrodynamics LLC
Recall StatusTerminated
Product CodeMOS
Date Initiated2019-08-20
Date Terminated2020-05-26
Product Quantity5 units
Reason for Recall
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
Firm Action
The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please
identify your affected customers and notify them.
3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide distribution - US in the states: NC OUS: Germany
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.