Xoran Mini CT Scanner
FDA Recall Z-0036-2014 · Posted None
Advertisement
Recalling FirmXoran Technologies, Inc.
Recall StatusTerminated
Product CodeJAK
Date Initiated2010-07-27
Date Terminated2014-09-11
Product Quantity320
Reason for Recall
A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu
Firm Action
All of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Worldwide Distribution - US nationwide