Medtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implantable Neurostimulator for Urinary Control.

FDA Recall Z-0037-2008 · Posted 2007-10-11

Recall StatusTerminated
Product CodeEZW
Date Initiated2007-06-25
Date Terminated2010-10-08
Product Quantity1885

Reason for Recall

A software issue may cause a lower than expected battery capacity estimate when evaluating an InterStim II Implantable Neurostimulator (model 3058) with an N'Vision Clinician Programmer(Model 8840) before implant .

Firm Action

Begining on July 11, 2007 a letter to the physicians was sent informing them of the issue and patient mangement of the two issues. The letter includes a Cover Letter, A Reply Form, and two Recall letters explaining the issue in detail.

Root Cause

Software design

Distribution Pattern

Nationwide and the Netherlands.