AGI-2237 Portable 2 Cylinder Manifold Cart, for connecting medical gas cylinders.
FDA Recall Z-0037-2010 · Posted 2009-10-14
Advertisement
Recalling FirmAirgas Puritan Medical
Recall StatusTerminated
Product CodeCAN
Date Initiated2009-08-06
Date Terminated2010-01-26
Product Quantity208
Reason for Recall
Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.
Firm Action
Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.
Root Cause
Component design/selection
Distribution Pattern
Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.