Terminated

AGI-2237 Portable 2 Cylinder Manifold Cart, for connecting medical gas cylinders.

FDA Recall Z-0037-2010 · Posted 2009-10-14

Recall StatusTerminated
Product CodeCAN
Date Initiated2009-08-06
Date Terminated2010-01-26
Product Quantity208

Reason for Recall

Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.

Firm Action

Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.

Root Cause

Component design/selection

Distribution Pattern

Nationwide distribution.

Official Source

View FDA Recall →