Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Datex-Ohmeda S/5 Aespire Anesthesia System 7900

FDA Recall Z-0038-06 · Posted 2005-10-12

Recalling FirmDatex-Ohmeda Inc
Recall StatusTerminated
Product CodeBSZ
Date Initiated2005-09-27
Date Terminated2007-04-01
Product Quantity16

Reason for Recall

Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.

Firm Action

Urgent Medical Device Correction letter dated 09/23/05 was sent to all customers. The letter describes the issues, affected device serial numbers and recommends that customers DO NOT USE the internal auxiliary oxygen flowmeter, if installed in their system, and to use alternate auxiliary oxygen sources, i.e. wall-mounted oxygen flow meter, in the interim. A GE service representative will contact the customers to schedule an upgrade, performed at no charge.

Root Cause

Other

Distribution Pattern

CA, CO, FL, GA, IL, IN, MI, MS, NM, OK, PA, PR, SD, UT & WA. OUS to: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Canada, Caribbean, Czech Republic, Chile, China, Colombia, Dominican Republic, Ecuador, Finland, France, Germany, Ghana, Great Britain, Hong Kong, India, Indonesia, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Oman, Palestine, Peru, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Arab Emerites, Venezuela & Vietnam.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑