Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.
FDA Recall Z-0038-2008 · Posted 2007-10-11
Advertisement
Recalling FirmRadiometer America Inc
Recall StatusTerminated
Product CodeCHL
Date Initiated2007-06-22
Date Terminated2012-04-25
Product Quantity289 pieces
Reason for Recall
Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur during the installation of the tubes or after a short time in use.
Firm Action
By telephone call or service representative visit beginning July 18, 2007, Consumers/End Users are asked to locate the analyzers and yearly service kits concerned and replace all pump tubes with new ones.
Root Cause
Process design
Distribution Pattern
Product was distributed throughout the United States and Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.