Terminated

RECLAIM Distal Reamer Extension Non Sterile REF 2975-00-500 Intended as an attachment device to the distal reamers to allow the surgeon to clear soft tissue while reaming the distal femoral canal.

FDA Recall Z-0038-2015 · Posted 2014-10-08

Recall StatusTerminated
Product CodeLXH
Date Initiated2014-09-03
Date Terminated2015-11-09
Product Quantity875 units

Reason for Recall

Specific lots are being recalled due to the potential for the tabs to break, leaving fragments in the patient.

Firm Action

On February 25, 2013, DePuy Orthopaedics, Inc. issued a device correction for the specific lots of the RECLAIM Distal Reamer Extension impacted by this recall. The letter stated the reason for the recall and swap out. A subsequent letter outlined a formal swap-out process of the affected reamer extensions as part of the device correction strategy. Newly designed devices are now available. Questions about device swap-out information may be directed to Kim Earle, Recall Coordinator, 574-371-4917; and Surgeon questions may be directed to DePuy's Scientific Information Office at 1-888-554-2482.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- USA, including the states of ME, MD, PA, MA, FL, LA, IL, IA, OH, MN, TX, MT, WA, CA, HI, NV, AZ, KY, WI, CO, GA, MI, NC, VA, NY, TN, and IN; and, the countries of Canada, Australia, Austria, Belgium, Czech Republic, United Kingdom, France, Germany, Ireland, Israel, New Zealand, Norway, Poland, Slovenia, South Africa, Switzerland, and Chile.

Official Source

View FDA Recall →