Coulter AcT 5diff Cap Pierce Hematology Analyzers Part Numbers: 6605641, 6605705

FDA Recall Z-0040-04 · Posted 2003-10-15

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2003-09-08
Date Terminated2004-03-30
Product Quantity451

Reason for Recall

Samples with high fat content will not count platelets accurately.

Firm Action

Firm issued an advisory letter to users to retain as part of the Quality Manual for the instruments. It cautions to check specimens for interfering lipemia. Letter was sent 9/8/2003.

Root Cause

Other

Distribution Pattern

United States and Canada