ATO Drill Guide Kit (All Through One Drill Guides Set), model number 07.00860.001, product is packaged in trays.
FDA Recall Z-0040-05 · Posted 2004-10-23
Advertisement
Recalling FirmZimmer Spine, Inc.
Recall StatusTerminated
Product CodeKWQ
Date Initiated2004-08-24
Date Terminated2005-11-03
Product Quantity82 sets
Reason for Recall
To date, there has been one complaint reported. During a Trinica surgical case, the surgeon was provided a standard disposable, Trinica Drill Bit (part number 07.00166.001) for use with the Trinica All Through One Drill Guide Instrumentation, instead of the ALL Through One Drill Bit. This enabled the physician to drill a few millimeters further than expected.
Firm Action
The product distributed to accounts will be returned to Zimmer Spine. Product will be quarantined pending investigation and determination of appropriate corrective action.
Root Cause
Other
Distribution Pattern
Product was distributed nationwide throughout the United States. Product was not distributed outside of the United States.