Xoran MiniCAT00202
FDA Recall Z-0040-2014 · Posted None
Advertisement
Recalling FirmXoran Technologies, Inc.
Recall StatusTerminated
Product CodeJAK
Date Initiated2010-07-27
Date Terminated2014-09-11
Product Quantity320
Reason for Recall
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.
Firm Action
All of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component. Xoran's Research & Development group is evaluating the current specification to see if the field specifications are appropriate and to ensure that proper testing protocols are communicated and are in place for use by 3 rd party testing organizations. In addition, Xoran has initiated CAPA # 18 0 that is looking at the source block with regard to evaluating the stability of the dose rating.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Worldwide Distribution - Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement