Terminated

McKesson Cardiology" Hemo intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

FDA Recall Z-0042-2015 · Posted 2014-10-08

Recall StatusTerminated
Product CodeMWI
Date Initiated2014-08-27
Date Terminated2016-05-18
Product Quantity272 devices

Reason for Recall

Software Error: The McKesson Cardiology Hemo calculation section incorrectly converts the Hemoglobin value before it is utilized in the applicable formula calculations.

Firm Action

On August 27, 2014, McKesson initiated a correction via a Field Safety Notice to all affected domestic customers. McKesson Israel will be responsible for contacting foreign consignees.

Root Cause

Software design (manufacturing process)

Distribution Pattern

Worldwide distribution: US (nationwide) including states of: AR, CA, FL, DE, GA, HI, IL, IN, KS, LA, MA, MD, MN, MO, MS, MT, NJ, NV, OK, PA, SC, TNTX, , UT, WA; and countries of: Australia, Canada, Israel and the United Kingdom.