Terminated

Bioplate Resorbable Bone Fixation Tack, Catalog Number: 89-0120, Item Number: SG#-45

FDA Recall Z-0043-2007 · Posted 2006-10-19

Recalling FirmBioplate Inc
Recall StatusTerminated
Product CodeJEY
Date Initiated2006-09-12
Date Terminated2011-09-14
Product Quantity36 units

Reason for Recall

Lack of device sterilization: An internal investigation found products were released for distribution labeled as sterile, but the lot was not sterilized.

Firm Action

Customers were notified via telephone initially on 09/14/2006 to return product, and this was followed up within two days with a certified mail/return receipt required, recall notification letter. 100% of the consignees will be notified of the recall. Bioplate will fax and call non-responsive customers afler 5 days via telephone call. Customers not responding to the first recall notification will be sent a second notification certified mail recall notification.

Root Cause

Other

Distribution Pattern

Nationwide to: CA, DC, GA, MI, OH, & PA