ESA LeadCare II Blood Test Kit, 6 test Kit containing Level 1 and Level 2 Control Vials with lot number: 0704A Catalog Number: 70-7142
FDA Recall Z-0043-2008 · Posted 2007-10-18
Advertisement
Recalling FirmEsa Biosciences, Inc.
Recall StatusTerminated
Product CodeDOF
Date Initiated2007-09-20
Date Terminated2008-03-25
Product Quantity2 kits
Reason for Recall
Level 1 and Level 2 Control values were incorrectly assigned.
Firm Action
ESA notifed Distributors and End Users by telephone and letter on 9/20/07. Distributors and End Users will be informed of the Level 1 and 2 Control value assignment change. End Users will be sent a new Level 1 and 2 Control Vial
label for each vial they originally received. Distributors will be asked to return all test kits in their inventory to ESA Biosciences.
Root Cause
Process design
Distribution Pattern
Nationwide and United Kingdom.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.