Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W ZTA2-P-42-173 ZTA2-P-42-173-W ZTA2-P-42-225 ZTA2-P-42-225-W ZTA2-P-44-125 ZTA2-P-44-125-W ZTA2-P-44-179 ZTA2-P-44-179-W ZTA2-P-44-233 ZTA2-P-44-233-W ZTA2-P-46-125 ZTA2-P-46-125-W ZTA2-P-46-179 ZTA2-P-46-179-W ZTA2-P-46-233-W ZTA2-PT-40-36-167 ZTA2-PT-40-36-167-W ZTA2-PT-40-36-217 ZTA2-PT-40-36-217-W ZTA2-PT-42-38-173 ZTA2-PT-42-38-173-W ZTA2-PT-42-38-225 ZTA2-PT-42-38-225-W ZTA2-PT-44-40-179 ZTA2-PT-44-40-179-W ZTA2-PT-44-40-233-W ZTA2-PT-46-42-179 ZTA2-PT-46-42-179-W ZTA2-PT-46-42-233 ZTA2-PT-46-42-233-W ZTA2-P-46-233 ZTA2-PT-44-40-233 ZTA2-P-40-217 ZTA2-P-40-117 ZTA2-P-40-167-DEMO ZTA2-P-44-152-W ZTA2-P-40-142 ZTA2-P-44-152 ZTA2-P-40-142-W

FDA Recall Z-0043-2026 · Posted 2025-10-10

Recall StatusOpen, Classified
Product CodeMIH
Date Initiated2025-09-18
Product Quantity695

Reason for Recall

Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization.

Firm Action

On September 18, 2025, the recalling firm sent Urgent Medical Device Recall letters to all affected customers. Customers were instructed to identify and quarantine unused affected product in inventory. Immediately cease all further distribution and use of the affected products. Affected product is to be returned to Cook Medical.

Root Cause

Process design

Distribution Pattern

Domestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC NJ NY OH OR PA SC TN TX UT VA WA WI WV International distribution to Switzerland, Germany, Italy, Poland, and Sweden

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑