Advanced Bionics implantable cochlear stimulator CLARION CII Models AB-5100H-11A AB-5100h-12A
FDA Recall Z-0045-05 · Posted 2004-10-23
Advertisement
Recalling FirmAdvanced Bionics
Recall StatusTerminated
Product CodeMCM
Date Initiated2004-09-27
Date Terminated2005-07-13
Reason for Recall
All unimplanted CLARION and HiResolution cochlear implants due to the potential presence of moisture in the internal circuitry, which can cause the device to stop functioning.
Firm Action
Expanded recall was initiated with notification to clinicians on 9/27/2004. Expanded effort includes: notice to patients implanted, notice to clinicians, worldwide return of all unimplanted coclear implants.
Root Cause
Other
Distribution Pattern
U.S. and Canada and France
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement