PERPOS PLS System, Single Use; CATALOG #'s 9045-01 & 9045-02. Interventional Spine, Inc. A sterile kit used to stabilize the spine as an aid to fusion through immobilization of the facet joints.

FDA Recall Z-0045-2012 · Posted 2011-10-13

Recall StatusTerminated
Product CodeHXX
Date Initiated2011-02-03
Date Terminated2013-02-05
Product Quantity2,620 units

Reason for Recall

This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result in a delay of the surgical procedure

Firm Action

Interventional Spine, Inc. sent an "Urgent Recall" letter dated March 15, 2011 to all affected customers. The letter included; affected product, reason for recall and recommendations. Customers were asked to immediately examine inventory, quarantine affected product and to notify International Spine, Inc if any product remain. For information on this recall contact the firm at (949) 472-0006.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Word wide Distribution: (USA) Nationwide including the states of CO, CT, FL, KY, LA, MA, NC, NY, OK, OR, TN, TX, VA, and WI; and the countries of Italy, Spain, Turkey, and the U.K.