Terminated

Bivona Adult Tracheostomy Tube, Labeled I.D. 6.0 mm, O.D. 8.7 mm, A 7.0 mm, B 0 mm, C 53.0 mm, Mfrd by Smiths Medical Critical Care; REF 60A160.

FDA Recall Z-0046-2007 · Posted 2006-10-19

Recall StatusTerminated
Product CodeJOH
Date Initiated2006-08-11
Date Terminated2007-12-04
Product Quantity71 units.

Reason for Recall

The length of the shaft is incorrectly stated as 7 mm, instead of the correct 17 mm, on the outer packaging, although the tray label is correct.

Firm Action

Consignees were notified to cease using, and to return the product, by recall letter dated August 2006 and mailed on 8/11/06.

Root Cause

Other

Distribution Pattern

Nationwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.