Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

FDA Recall Z-0046-2018 · Posted None

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeJJE
Date Initiated2017-08-03
Date Terminated2023-09-27
Product Quantity27 units

Reason for Recall

Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.

Firm Action

BEC will initiate a SW update- to fix the SW defect. This will be executed via MOD (field modification) performed by Field Service.

Root Cause

Software design

Distribution Pattern

China France Italy Kuwait Spain Turkey United States