Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4.1 (excluding version 6.3.8) Common name: Medical image storage device. Class I, 510 (k) exempt. (NOTE: any ImageManager Software sold as a stand-alone system is not impacted by this software malfunction since no opportunity exists for any image cross-linking. It is a condition of the software malfunction that the observed image crosslink results from the collision of data files from two separate ImageManager computers.)

FDA Recall Z-0047-04 · Posted 2003-10-15

Recalling FirmOlympus America Inc.
Recall StatusTerminated
Product CodeLMB
Date Initiated2003-08-19
Date Terminated2005-03-22
Product Quantity1014

Reason for Recall

Software malfunction; potential for simultaneous cross-linking of exam images to another patient's file.

Firm Action

OAI sent "Urgent-Device Correction" letters, dated 8/19/03, to the end users, explaining the software malfunction. The letter also provided specific instructions to the end users until the software fix is installed. For technical assistance, users can contact OAI's Technical Assistance Center 1-800-848-9024. Accompanying the letter is a questionnaire to be completed by the end users & returned to OAI for scheduling an appointment to visit the facility to install the software fix.

Root Cause

Other

Distribution Pattern

Medical facilities (end users) nationwide and Puerto Rico, plus 5 foreign accounts located in Canada, Mexico, Ecuador and Chile.