Recall StatusTerminated
Date Initiated2006-08-08
Date Terminated2010-01-08
Product Quantity950
Reason for Recall
This action is being initiated due to power supply snubber board failures on certain Esprit Ventilators in countries that utilize operating voltages of 200-240 Volts AC and will involve the replacement of
the Esprit Ventilator power supply snubber board. Returned power supplies have shown evidence of overheating and in some cases in which a fan also failed, have shown signs of fire damage.
Firm Action
The field action will be conducted through consignees (distributors) in the affected countries. The distributors were notified of the field action by Federal Express au on August 8th, 2006. The distributors have been informed that Respironics will provide replacement circuit boards at no cost and will be instructed to upgrade the ventilator power supplies in their region. Instructions have been provided to the distributors and users on how to evaluate the ventilator to be able to continue to use it pending the completion of the upgrades.
The distributors will document the replacement of the affected component by signing and dating the activity on the Ventilator Field Action Log (reference sample on Appendix IV). The serial number of the power supply destroyed will also be recorded on the log.
Distribution Pattern
Worldwide, including Algeria,
Argentina,
Australia,
Bangladesh,
Brazil,
Cambodia,
Chile,
Egypt,
France,
Hong Kong,
India,
Italy,
Jordan,
Lebanon,
Malaysia,
Pakistan,
Peru,
Philippines,
Russia,
S Africa,
S Korea,
Spain,
Thailand,
Turkey,
United Kingdon, &
Vietnam
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.