Terminated

Fabius GS Anesthesia Machine. Catalog number 8604699

FDA Recall Z-0048-2007 · Posted 2006-10-19

Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeBSZ
Date Initiated2006-10-03
Date Terminated2007-08-07
Product Quantity429 units

Reason for Recall

One of four casters may break loose from chassis

Firm Action

The recalling firm issued recall letters to their customers 10/03/2006 to inform them of the problem and that a service represenative will contact them to schedule an inspection of their machines. Should the inspections revealed any affected machines, the machines will be repaired free of charge.

Root Cause

Other

Distribution Pattern

Nationwide and Canada.