Siemens syngo.plaza radiological image processing system.

FDA Recall Z-0048-2013 · Posted 2012-10-11

Recall StatusTerminated
Product CodeLLZ
Date Initiated2012-06-06
Date Terminated2013-04-13
Product Quantity11

Reason for Recall

Siemens became aware of an unintended behavior when using syngo.plaza. In syngo.plaza within the online "Short Time Storage" (STS), some studies are marked with an archived icon, however, the entry in the Archive Database referring to the physical file may be missing.

Firm Action

The firm initiated this correction on June 6, 2012 by issuing update instructions SY129/12/S to consignees. A script will be distributed to all potential affected customer systems which will prevent data loss by setting the delete protection on all affected studies.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution including the states of MO, FL, MA, NY, WI, GA, CA and WA.