STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.
FDA Recall Z-0048-2014 · Posted 2013-10-22
Recalling FirmAbbott Medical Optics, Inc.
Recall StatusTerminated
Product CodeLZS
Date Initiated2013-09-16
Date Terminated2014-11-12
Product Quantity834 units in total
Reason for Recall
This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).
Firm Action
The firm, Abbott Medical Optics, Inc., (AMO), sent an "ADVISORY NOTICE and a COVER LETTER FOR ADVISORY NOTICE" dated September 16, 2013 to all affected customers. The letters identified the product, problem, and actions to be taken by the customers. The customers were instructed to complete and return the Star Notification Acknowledgement form via fax to 1-714-247-4530 within 3 business days or mail to Abbott Medical Optics, 510 Cottonwood Drive, MS 347118, Milpitas, CA 95035, and share this advisory notice with your staff.
If you have any questions related to operation of STAR Excimer Laser System or documentation requirements associated with this letter, please call AMO Technical Support at 1-800-511-0911 and you will be directed to an AMO Representative.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.