GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator
FDA Recall Z-0048-2025 · Posted 2024-10-17
Advertisement
Recalling FirmDATEX--OHMEDA, INC.
Recall StatusOpen, Classified
Product CodeFMZ
Date Initiated2024-09-03
Reason for Recall
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
Firm Action
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/03/2024 via Fed-Ex. The notice explained the problem with the device, safety issue, and requested the following:
Download the Service Manual Addendum (5971733) from the GE HealthCare Customer
Documentation Portal: https://www.gehealthcare.com/support/manuals.
Inspect the heater door mounting screw on your device. If the screw is not appropriately tightened, complete steps 9 and 10 of Section 1.1 of the Service Manual Addendum (5971733).
Root Cause
Use error
Distribution Pattern
Worldwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.