DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-1330; Phantom model (PH-150-CM & PH-150-G)
FDA Recall Z-0049-2007 · Posted 2006-10-19
Recalling FirmDel Medical Systems Group
Recall StatusTerminated
Product CodeIZL
Date Initiated2006-10-03
Date Terminated2008-01-24
Product Quantity53 units
Reason for Recall
The x-ray tube head assembly of the Phantom Portable X-Ray System may detach from the boom arm assembly. The tube head could fall and contact the patient or operator.
Firm Action
Del Medical sent Field Correction Recall letters dated 10/3/06 to the direct accounts, informing them of the potential for the shoulder bolt in the yoke assembly to come loose. If the shoulder bolt comes loose, the tube head could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Root Cause
Other
Distribution Pattern
Nationwide and Thailand