DABRA RA-308 Excimer Laser, SN: ******* - Product Usage: The device is intended for use in ablating a channel in occlusive peripheral vascular disease.

FDA Recall Z-0050-2021 · Posted None

Recall StatusTerminated
Product CodePDU
Date Initiated2020-07-24
Date Terminated2023-08-09
Product Quantity54 units

Reason for Recall

Due to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury.

Firm Action

On August 10, 2020, the firm sent an "Urgent Medical Correction" letter to customers, informing them of the incorrect laser wheel size, which upon moving or adjusting the laser with the wheels in locked position may fall over. If this were to happen during a medical procedure, user or patient injury may occur.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, FL, ID, IL, IN, KS, KY, LA, MA, MD, MI, NC, NV, NY, OR, OK, PA, TX, WA and the country of Japan.