Kendall Monoject Syringe with Hypodermic Needle, 3cc syringe with 27 ga A-Bevel Needle (3cc- 27x1-1/4 A) Product # 8881513744

FDA Recall Z-0051-2007 · Posted 2006-10-21

Recall StatusTerminated
Product CodeFMF
Date Initiated2006-09-11
Date Terminated2009-09-11
Product Quantity130,000

Reason for Recall

Incorrect needle: There is an incorrect needle configuration on the syringe. The correct needle has a 27 gage A-bevel, the needle on the syringe has no bevel, typically used for dental irrigation.

Firm Action

A recall letter was mailed 9/12/2006 to the 44 customers who purchased this product from Tyco Healthcare. The firm requested that their distributors notify their customers, and coordinate recovery of the product. The recall mailing included a form that was to be faxed back to Tyco Healthcare. A postage paid return envelope was also included if the customer does not have access to a fax machine.

Root Cause

Other

Distribution Pattern

Nationwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →