Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

K2M Special Connector Instruments (Special Connector, Special Adaptor, and Quick Adaptor) Product Usage: Special Connector Instruments include an additional feature on their shafts that enable surgeon users to combine them with non-K2M navigation systems as part of their standard surgical technique.

FDA Recall Z-0052-2020 · Posted None

Recalling FirmK2M, Inc
Recall StatusTerminated
Product CodeOLO
Date Initiated2016-12-02
Date Terminated2020-05-22
Product Quantity503

Reason for Recall

Instruments were developed without adequate consideration given to the need for a new regulatory filing. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; recall actions were completed in 2017.

Firm Action

K2M notified customers with a Recall Notification Letter dated December 2, 2016. The letter instructed customers to locate and segregate recalled product, complete a tracking/verification form, and return the product to K2M. The product was reintroduced to the market after K2M obtained new 510(k) clearance for the product. for question call at 571-919-2000

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide Distribution in the states of AL CA CO CT FL FL IL IN KS MA MI MN PA TN TX UT VA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑