SARNS INTRACARDIAC SUCKER, BULK, NON-STERILE
FDA Recall Z-0054-2012 · Posted 2011-10-17
Advertisement
Recalling FirmTerumo Cardiovascular Systems Corporation
Recall StatusTerminated
Product CodeDTS
Date Initiated2011-09-13
Date Terminated2012-12-10
Product Quantity5700
Reason for Recall
Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the stainless steel tube of the Sarns" Adult Rigid Intracardiac Sucker.
Firm Action
On 9/13/2011 Terumo sent a letter entitled "URGENT MEDICAL DEVICE REMOVAL" to its customers that included the name and description of the device, lot numbers, the potential hazard, correction, the reason for the recall and instructions. Customers are to review the notice, ensure that all users are aware of the issue, and a recall response form to be completed and returned. Customers are to call Terumo CVS Customer Service to obtain a Return Goods Authorization number. The letter also included MedWatch reporting information and stated that Customers with questions should call 1-800-521-2818.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution -- USA, Belgium, Australia, Mexico, and Japan.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.