VERIGENE Processor SP Model No. 10-0000-07. An in vitro diagnostic device intended for processing and identifying nucleic acids
FDA Recall Z-0054-2021 · Posted None
Advertisement
Recalling FirmLuminex Corporation
Recall StatusOpen, Classified
Product CodeOCC
Date Initiated2018-12-10
Product Quantity1 system
Reason for Recall
No results/incorrect results due to failure of the hybridization heater.
Firm Action
On or about 11/19/2018 there was one complaint from a customer regarding a thermal control failure. In response to the complaint the Field Service Engineer replaced hybridization heater and ran all tests as passed. Subsequently, the customer reported the initial service was ineffective. The VERIGENE SP Processor was then removed from the customer site and returned for analysis.
No Customer Notification was communicated for this removal.
Root Cause
Component design/selection
Distribution Pattern
US distribution in MO. No OUS