Terminated

8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable Port is a Single Titanium Port System with Attachable 8.0 Fr. Polyurethane Catheters and 8Fr. Introducer Kit.

FDA Recall Z-0056-2012 · Posted 2011-10-13

Recalling FirmAngiodynamics, INC
Recall StatusTerminated
Product CodeDYB
Date Initiated2011-08-01
Date Terminated2012-03-07
Product Quantity1095 total units

Reason for Recall

The product may contain an incorrect size catheter and introducer.

Firm Action

AngioDynamics, Inc contacted affected customers initially via telephone and a follow up "Notification of Medical Device Correction" letter was sent dated August 1, 2011. The letter informed them of the problem and provided instructions. Customers were instructed to utilize the 8F Introducer Sheath provided within the 8F Smart Port CT Implantable Port during port implantation. The 8F Valved Sheath (V18) may continue to be utilized during the placement of the 7.5F Smart Port Implantable Port (for example: CT75STSD; CT75STSD-NF). For question on this recall contact AngioDynamics, Inc at (800) 772-6446.

Root Cause

Device Design

Distribution Pattern

(USA) Nationwide distribution including the states of CA, FL, GA, IA, IL, IN, KY, MD, MI, MO, NC, NE, NJ, NY, OH, OK, OR, TN, UT, WA and WI.