Terminated

Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IP2.P Rx Only, Manufacturer GMS- Gesellschaft fur medizinische, Sondentechnik mbH Dofstrasse 2 24247 Melkendorf, Germany Distributed by Integra Neurosciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Used in direct monitoring of the partial pressure of oxygen.

FDA Recall Z-0056-2014 · Posted 2013-10-23

Recall StatusTerminated
Product CodeGWM
Date Initiated2013-06-11
Date Terminated2014-02-18
Product Quantity503 kits(482 US; 21 non US)

Reason for Recall

Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.

Firm Action

Integra LifeSciences Corporation sent a recall letter/return response form on 6/11/2013 by traceable courier service, traceable e-mail or facsimile.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution: US (nationwide) and countries of: AR, BR, CH, CL,CR. DE, ES, FI, FR, GB, IL, IT, MX, PA, PL, PT, RU, TW, and ZA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.