Terminated

Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and 3.1.x - Product Usage: intended to be an intra-operative image guided localization system to enable minimally invasive surgery.

FDA Recall Z-0056-2021 · Posted None

Recalling FirmBrainlab AG
Recall StatusTerminated
Product CodeHAW
Date Initiated2020-08-31
Date Terminated2021-02-17
Product Quantity271 units

Reason for Recall

Brainlab Ultrasound Navigation Software does not support the modification of the probe's image width.

Firm Action

The intraoperative Brainlab Ultrasound Navigation Software does not support the modification of the probes image width, an optional feature of the BK Medical Ultrasound System, resulting in an incorrect display of the depth of the ultrasound overlay by the navigation for non-linear ultrasound probes, with the deviation depending on the probe and the difference between the calibrated and used image width.

Root Cause

Software design

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of Illinois, Massachusetts, Texas, California, Idaho, Missouri, Oregon, Connecticut and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Curacao, Czech Republic, Egypt, Finland, France, Germany, Switzerland, Netherlands, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kuwait, Malaysia, Myanmar, Norway, Pakistan, Romania, Slovakia, South Africa, Spain, Sweden, Thailand, Turkey and United Kingdom.