Depuy PFC Oval Dome Patella (Part Number 960100)3-Peg, 32 mm, sterile; REF 96-0100.
FDA Recall Z-0057-2007 · Posted 2006-10-21
Advertisement
Recalling FirmDepuy Orthopaedics, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2006-09-14
Date Terminated2007-04-26
Product Quantity9
Reason for Recall
Mislabeled: units containing a size 32 mm. patella were labeled as a 38 mm patella.
Firm Action
The firm called their sales staff on 9/14/06 and instructed them to place a hold on the product, which entails pulling products from their own inventory and at any hospitals where it has been placed on consignment. Sales staff immediately visited hospital accounts and switched out the recalled lot for other product.
Root Cause
Other
Distribution Pattern
Nationwide.