ON-Q PainBuster with OnDemand (270 ml, 2 ml/hr + 5 ml bolus /60 min refill), Model PMB01, I-Flow Corporation, Lake Forest, CA 92630 U.S.A.
FDA Recall Z-0057-2008 · Posted 2007-10-06
Advertisement
Recalling FirmI-Flow Corporation
Recall StatusTerminated
Product CodeMED
Date Initiated2007-03-27
Date Terminated2009-10-07
Product Quantity2,587 units
Reason for Recall
Misassembly: The ON-Q PainBuster PMB01 package may contain a Soaker catheter instead of a Non-Soaker catheter. As a result, the bolus button may remain latched in the down position which may cause the the device to deliver continuously at 7 ml/hr instead of 2 ml/hr.
Firm Action
An Important Recall Notification Letter was faxed to affected customers on March 27, 2007. Affected customers were notified of the recall, the reason for recall, and instructed to immediately quarantine any remaining inventory. A Customer Response (and Fax back) form was included with the Recall Notification. Affected customers were requested to return the FAX back form to I-Flow within 48 hours of receipt. Replacement product will be sent out as applicable.
Root Cause
Process control
Distribution Pattern
Worldwide, USA and Australia