HeartMate II System Controller, a component of the HeartMate II Left Ventricular Device System (LVAS), distributed both as stand alone devices and as components of the LVAS Implant kit. Manufactured/Distributed by Thoratec Corporation, Pleasanton, CA The HeartMate II Left Ventricular Assist System (LVAS) consists of an implantable blood pump connected to an eternal system controller by a percutaneous lead. The external controller is powered bye either batteries or a power supply that connects to AC Main power. The device is intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. It is also for use in patients with New York Heart Association class IIIB or IV end-stage left ventricular

FDA Recall Z-0058-2011 · Posted 2010-10-13

Recalling FirmThoratec Corp
Recall StatusTerminated
Product CodeDSQ
Date Initiated2010-09-01
Date Terminated2011-04-07
Product Quantity8965 units

Reason for Recall

Connection failure-- product may malfunction because of bent pins or broken wires in power cord connection

Firm Action

Thoratec Corporation sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 30, 2010, to all customers. The letter described the affected product, description and symptoms of problem, and immediate action to be taken by the customers. The customers were instructed to: 1) contact all ongoing patients using the affected Systems Controllers (serial numbers starting with "EPC") to determine if they are experiencing any unexplained intermittent alarms or have any bent pins in the connectors between the System Controller and the power sources, if so, have them return to the hospital so that their power leads can be examined. If further assistance is needed with the evaluation, please contact Thoratec Technical Service (800-456-1477). 2) review the attached addendum to the HeartMate II labeling with all VAD clinicians and ongoing HeartMate II LVAS patients, and 3) complete and sign the attached Acknowledgment Form via fax to Thoratec Regulatory Affairs at (952) 847-8571. If you have any questions, please contact 925-847-8600.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution: USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX , UT, VA, WA, and WI; and countries including: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Netherlands, Singapore, Slovenia, Sweden, Switzerland, and the UK.