AIR Technology Anterior Array, Catalogue Numbers: (a) 5746694, (b) 5746695, (c) 5746696, a receive-only RF coil designed for use with GE 3.0T MRI systems.

FDA Recall Z-0062-2020 · Posted None

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeMOS
Date Initiated2019-03-19
Date Terminated2020-03-30
Product Quantity22 units

Reason for Recall

The outer seam of the coil, closest to the system cable, may separate and expose the porous material internal to the coil. This could result in user and/or patient contact with infectious agents since the internal porous material cannot be disinfected.

Firm Action

The firm hand delivered the notices beginning on 03/19/2019. The notices described the separation of the outer seam and requested that the consignee check each coil component for tears or damage, and to discontinue use of product if product is suspected of not being in good condition. The affected products will be replaced. For questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of AZ, CA, MS, NY, WI, DE, and countries of France, Germany, Japan, Netherlands, Spain, Sweden,