Pagewriter Touch Cardiograph Software Versions A.01.01 or A.01.02

FDA Recall Z-0063-05 · Posted 2004-11-02

Recall StatusTerminated
Product CodeDPS
Date Initiated2004-10-13
Date Terminated2012-05-15
Product Quantity768 units

Reason for Recall

Software can generate printed ECG's that associate incorrect patient data with the waveform

Firm Action

Philips notified accounts by letter on 9/13/04 and advised users of the problem and will provide the software upgrade kit to version A.01.03.

Root Cause

Other

Distribution Pattern

Nationwide Australia, Brazil Korea, Japan, Mexico, Singapore,Taiwan,Canada, Belgium, Germany, Lebanon, Israel, Italy, Ireland, Netherlands, Oman, Spain, Saudia Arabia,Poland, Qtar,Switzerland, UAE, UK,