Tru-Link DIN unshielded ECG cable, 5-lead
FDA Recall Z-0063-06 · Posted 2005-11-03
Advertisement
Recalling FirmSpacelabs Medical Incorporated
Recall StatusTerminated
Product CodeDSI
Date Initiated2005-09-16
Date Terminated2006-02-14
Product Quantity632 cables
Reason for Recall
ECG 5-lead cable miswired - A physician could interpret the incorrect polarities as a conduction defect.
Firm Action
On 9/16/05 the firm sent a letter to their customers advising of the problem. The customer was requested to check the code on their cables, and if the suspect code, discard the cable. The customer was provided a response sheet to indicate they received the notification and to request new cables.
Root Cause
Other
Distribution Pattern
Cables were distributed to 9 US hospitals/medical centers and 30 international customers.