Medtronic Export XT Aspiration Catheter, 6F (US) Medtronic Export XT Aspiration Catheter, CE, 6FR (Outside US)

FDA Recall Z-0063-2008 · Posted 2007-10-31

Recalling FirmMedtronic, Inc.
Recall StatusTerminated
Product CodeDQA
Date Initiated2007-08-29
Date Terminated2008-09-08
Product Quantity13,940 units

Reason for Recall

Additional Warning in Instructions for Use for guide wire management. Complaints of wire lumen tearing during use or band marker becoming dislodged.

Firm Action

Medtronic has notifed sales representatives to visit Cath lab site with the letter dated August 29, 2007 and educate the user, provide the Safety Alert Letter and document delivery of the letter with the confirmation form.

Root Cause

Labeling design

Distribution Pattern

Worldwide, including USA, Canada, India, Austria, Belgium, France, Germany, Italy, Netherlands, Spain, Sweden, Switzerland, and UK.