Blood Pressure Monitor UA-767PBT-Ci for small cuff. Designed to be used by end users who are eighteen (18) years and older at home to monitor their blood pressure (systolic and diastolic) and pulse rate.

FDA Recall Z-0063-2017 · Posted None

Recalling FirmA & D Enginering Inc
Recall StatusTerminated
Product CodeDXN
Date Initiated2016-09-12
Date Terminated2018-08-28
Product Quantity378 units

Reason for Recall

A&D may have shipped blood pressure monitors that were not labeled for use with small cuff.

Firm Action

Notification letters were sent on September 12, 2016 to all affected customers. Letters identify the product and reason for the recall. Letters include a serial number list that is customer specific and ask that inventory be reviewed and that other organizations that may be impacted by this issue be notified.

Root Cause

Labeling Change Control

Distribution Pattern

to distributors in NY, MA, VA. TX,