Terminated

Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.

FDA Recall Z-0064-2019 · Posted None

Recall StatusTerminated
Product CodeKPA
Date Initiated2018-06-13
Date Terminated2019-03-18
Product Quantity67 units

Reason for Recall

These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.

Firm Action

On approximately 06/13/2018 sent a Field Action Notices to consignees. Instructions included to make arrangements with their Leica Service Engineer to correct affected devices, ensure all appropriate personnel are made aware of the recall, and complete and return the Field Action Notice Return Response Form to confirm receipt and understanding of the recall notice.

Root Cause

Device Design

Distribution Pattern

Distributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI. Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND