BD Microkeratome K3000 and K4000;Catalog Number: 378660, Becton Dickinson & Company (BD)
FDA Recall Z-0066-2008 · Posted 2007-10-23
Advertisement
Recalling FirmBecton Dickinson and Company
Recall StatusTerminated
Product CodeHMY
Date Initiated2007-08-09
Date Terminated2012-02-29
Product Quantity256 units
Reason for Recall
Microkeratome blade may stop oscillating and transverse the eye causing injury to the cornea
Firm Action
BD Medical notified customers by Advisory Letter dated 8/09/07, advising users to return heads and bearing cartridges for replacement of the o-ring.
Root Cause
Component design/selection
Distribution Pattern
Worldwide