Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

AC Powered adjustable bed frame. AC powered adjustable bed frame used in nursing facilities.

FDA Recall Z-0067-2015 · Posted 2014-10-15

Recalling FirmPrimus Medical LLC
Recall StatusTerminated
Product CodeFNL
Date Initiated2014-09-15
Date Terminated2017-08-10
Product Quantity1726

Reason for Recall

Firm received several complaints that the head deck actuator bracket had separated from the bed frame.

Firm Action

Primus Medical will send a letter notifying all customers with PCB900 beds at their facility of the potential head bracket separation.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of CA, DE, FL, GA, IL, IN, KS, KY, MA, MN,MS, NC, NM, NV, NY, OH, PA, TN, TX, and VA, and the country of Canada.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑