Delta Cup Beater Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System

FDA Recall Z-0068-2018 · Posted None

Recalling FirmLimacorporate S.p.A
Recall StatusCompleted
Product CodeLXH
Date Initiated2017-09-08
Product Quantity20 units in (US)

Reason for Recall

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

Firm Action

The Firm, Lima Corporate has notified the US agents of the recall and asked to check each single set with the affected product codes, identify the items and remove them from the inventory. The US distributors have also been instructed to fill out and sign a response document specifying the quantities returned. For further questions, please call (800) 962-2578.

Root Cause

Vendor change control

Distribution Pattern

Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany