Sienet Cosmos V15C, picture archiving and communications system. Model number 10018969
FDA Recall Z-0069-2007 · Posted 2006-10-25
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeLLZ
Date Initiated2006-10-03
Date Terminated2008-04-23
Product Quantity5 units
Reason for Recall
Potential loss of images. Images may not successfully be sent to the RAID system.
Firm Action
The recalling firm has issued a software patch to resolve the issue via Update Instructions HS110/06/S. The Update Instructions direct a Siemens Service Representative to visit the affected sites, install the software patch to resolve the issue, and hand deliver a Customer Advisory Notice, dated September 20th, 2006, to the customer to notify them of the issue.
Root Cause
Other
Distribution Pattern
The product was shipped to hospitals in CA, MD, MO, and PA