Recall StatusTerminated
Date Initiated2018-08-17
Date Terminated2020-04-10
Product Quantity1,902
Reason for Recall
The firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete staple formation and/or the inability to remove the device from tissue following application potentially leading to bleeding, anastomotic leak or tissue trauma. There have been no reports of serious injury related to this issue.
Firm Action
All U.S. consignees were notified via Federal Express or certified mail
on August 17, 2018. The letter informs customers of the potential
for use of a device to have an incorrect tissue gap.
Required Actions:
1. Please quarantine and discontinue use of the affected item code and lots listed above.
2. Please return affected product. All unused products from the affected item codes and lots must be returned. If you purchased directly from Medtronic, return all inventory of the affected lots to Medtronic (Covidien), 195 McDermott
Road, North Haven, CT 06473, Attention: Field Returns Department. If you purchased from a distributor, complete all fields on the form and contact your distributor directly to arrange for return of product.
3. If you have distributed EEA" staplers with Tri-Staple" technology listed above, please promptly forward the information from this letter to those recipients.
4. Complete the Recalled Product Return Form even if you do not have inventory.
If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of CA, FL, IL, NC, NY, UT. and the countries of Austria, Belgium, Denmark, Finland, France, Germany, Iceland, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Reunion, Spain, Sweden, Switzerland, and United Kingdom.