VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007
FDA Recall Z-0071-2027 · Posted 2026-10-02
Advertisement
Recalling FirmSteris Corporation
Recall StatusOpen, Classified
Product CodeFRC
Date Initiated2026-08-31
Product Quantity409
Reason for Recall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Firm Action
Firm began notifying customers on August 31, 2026 via URGENT MEDICAL DEVICE RECALL letters.
Customers were instructed to destroy any affected product on hand. The firm will replace any product still within its shelf life.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide - US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.