Terminated

Biotrak PLLA Pin, Resorbable Fixation System, Biotrak 20 mm Pin; sterile pin packaged in sterile bag, enclosed within a box labeled Biotrak Resorbable Fixation System Biotrak 20 mm Pin REF 30170320-S Manufactured or Distributed by: ACUMED LLC 5885 NW Cornelius Pass Rd Hillsboro, OR 97124. The Acumed resorbable fixation devices are designed to provide fixation of specific fractures, osteotomies, and arthrodeses while they heal.

FDA Recall Z-0073-2010 · Posted 2009-10-23

Recalling FirmAcumed LLC
Recall StatusTerminated
Product CodeHWC
Date Initiated2009-08-21
Date Terminated2010-11-06
Product Quantity125

Reason for Recall

Labeling may contain 2 lot numbers and 2 model numbers. It was observed that the box top label (30170060-S, W67391) does not match the two side labels (3010320-S, W67385). The part and patient labeling inside the box match the side labels (30170320-S, W67385).

Firm Action

Urgent Notice: Device Recall Letters dated August 21, 2009 were sent to all consignees advising of the issue and and required action. Customers are to identify and quarantine the product having the affected lot code(s). Customers are to also return all product from the specified lot(s) to Acumed immediately. If product was further distributed, then those consignees should be notified of the recall and return of the product(s) should be requested. Questions should be directed to Acumed Customer Service at 1-888-627-9957.

Root Cause

Pending

Distribution Pattern

Worldwide Distribution -- USA and South Africa.